Comparison Of Neutron Coincidence And Multiplicity Counting Techniques For Safeguards
Dorothy Davidson and Robert McElroy
CANBERRA Nuclear Products Group
Meriden, Connecticut, USA

Abstract

Neutron counters have been used by safeguard inspectors worldwide for over 20 years to

accurately perform Non-Destructive Assays of pure samples of plutonium and uranium. These systems rely on the neutron coincidence counting technique to separate time-correlated fission neutrons from random, uncorrelated neutrons to determine the fissile mass. Impure samples such as mixed-oxide (MOX) scrap materials and salts present a unique problem because they cannot be accurately measured using the standard coincidence technique. Instead, they require an enhanced technique, multiplicity counting, to accurately quantity the fissile content. Test data will be presented in this paper to compare multiplicity and coincidence counting, and to identify sample types that can be assayed using these techniques.

I. Introduction

Several different types of neutron counters that use traditional coincidence counting techniques have been authorized for routine inspection use (AFU) by the IAEA and Euratom, to make accurate verification measurements of pure PuO2 samples. Such AFU systems include the Inventory Sample Coincidence Counter (INVS), High Level Neutron Coincidence Counter (HLNC), and the Active Well Coincidence Counter (AWCC). Impure plutonium samples, or samples with unknown chemical composition require a more sophisticated technique, multiplicity, to accurately measure the plutonium content. This paper will briefly discuss the standard coincidence and multiplicity techniques, and review test data for MOX standards and scrap material.

II. Discussion of the Techniques

Traditional coincidence counting measures two parameters: the totals neutron rate and the reals or time-correlated rate. The reals rate is determined by subtracting the counts in the Reals+Accidentals (R+A) gate from the Accidentals (A) gate (Figure 1). The gate width for the R+A and A gates varies depending on the counter.


Figure 1.
Simplified diagram showing howthe counter response as a function of time.

With these two measured values, R+A and A, one can determine the plutonium mass and multiplication provided the ratio of random neutrons to coincidence neutrons (e.g. alpha) is known. The equation for alpha is

a 0

where the isotopic fraction for isotope i are represented by fi, and the values for ki are known constants for each isotope. For impure plutonium samples it is not valid to assume that the alpha value is known. An enhanced analysis technique, called multiplicity, is required to accurately assay the material.

The multiplicity counting technique has been under investigation at ISPRA and Los Alamos National Laboratory (LANL) for over 10 years to address the problem of NDA for impure samples or samples of unknown chemical composition. The uncertainty in the chemical form of the plutonium leads to an unpredictable level of (?,n) neutron emission. The analysis then requires solution for three unknown values: fissile mass, neutron multiplication, and alpha. Multiplicity counting provides a third measured value allowing a unique solution for the three unknowns. Impure plutonium samples which cannot be analyzed reliably with standard neutron coincidence counting can thus be analyzed with this technique.

Canberra’s 2150 multiplicity electronics module is an extension of the IAEA AFU JSR-12 Neutron Coincidence Counter Electronics (NCCE). Where the JSR-12 simply sums the R+A and the A gates, the 2150 multiplicity module records the number of times in 256 scalers that a given number of neutrons is recorded. The R+A and A multiplicity distributions for a MOX standard are listed in Table 1.

Multiplicity Reals+ Accidentals Counts Accidentals Counts Multiplicity Reals+ Accidentals Counts Accidentals Counts
0 9605 11373 20 1186436 1039227
1 83045 98659 21 716308 619723
2 375384 434229 22 416231 356067
3 1151735 1312457 23 237073 199777
4 2694830 3010470 24 130075 108311
5 5160160 5662859 25 69755 57013
6 8405588 9069565 26 36816 29950
7 11960226 12676951 27 18642 15155
8 15166519 15827083 28 9647 7555
9 17432091 17882545 29 4728 3710
10 18348968 18537609 30 2154 1637
11 17895079 17798142 31 1014 723
12 16281194 15961957 32 436 344
13 13920981 13436624 33 222 166
14 11235911 10689922 34 108 73
15 8610825 8084184 35 46 41
16 6293619 5824891 36 14 10
17 4405350 4013039 37 9 12
18 2956488 2660166 38 4 6
19 1906225 1691402 39 0 0

Table 1.
Multiplicity Distribution for 1000-sec Measurement of MOX Standard

The multiplicity distribution is used to determine the singles (S), doubles (D) and triples (T) rates as defined by the equations below.

neutron_drw_small.gif (3169 bytes)
Click for enlarged view. To return to this page, click the "Back" button.

where m is the 240Pu-effective mass, M the multiplication, e the efficiency, and fd and ft the doubles and triples gate fractions. These equations are explained in detail elsewhere.

III. Test Results

The Plutonium Scrap Multiplicity Counter (PSMC), shown in Figure 2, is a commercialized version of a multiplicity counter originally developed by Los Alamos National Laboratory, Group NIS-5, to assay MOX scrap. This PSMC was used to measure well characterized MOX standards of known chemical composition and plutonium isotopics to compare the known alpha technique (traditional coincidence counting) and the multiplicity technique.

Figure 2.
Plutonium Scrap Multiplicity Counter

The PSMC design is optimized for multiplicity counting. The important features include:

  • High efficiency - As shown in the equation above, the triples rate is proportional to the efficiency cubed. The PSMC has 80 3He proportional detectors to obtain an efficiency of ~55%.
  • Uniform Response- The PSMC has four rings of 3He proportional detectors arranged to flatten the energy response, and graphite end plugs to flatten the axial response by reflecting neutrons from the dead region of the end plugs back into the 3He detection region.
  • Small deadtime - The counter has 19 amplifiers to reduce the deadtime.

Performance characteristics of the PSMC are published in detail elsewhere.

The MULTI code1 uses both the known alpha technique and the multiplicity technique to analyze the same data set. Results for MOX standards are listed in Tables 2 and 3. Each standard was measured for 10 cycles, 100-sec each for a total of 1000 seconds. The final column in each table is the difference between the declared mass and the measured mass averaged for four 1000-sec assays. This difference gives the accuracy of each verification measurement.

Standard Totals Rate (error) Reals Rate (error) Multiplication Declared-Measured Mass (%)
1 259080.90 (± 16.3) 56255.19 (± 97.08) 1.07 -1.74
2 171745.28 (± 13.2) 36027.25 (± 63.49) 1.06 -0.37
3 80217.66 (± 8.99) 15441.09 (± 29.35) 1.04 0.78

Table 2.
Known Alpha Analysis of 1000-sec Measurements of MOX Standards

Standard Singles Rate (Error) Doubles Rate (Error) Triples Rate (Error) Alpha Multiplication Declare - Measured Mass (%)
1 255261.18 (± 25.01) 57178.12 (± 0.92) 35137.13 (± 10.28) 0.861 1.080 -0.12
2 170026.8 (± 14.89) 36484.95 (± 0.55) 19154.60 (± 10.98) 0.749 1.057 -1.47
3 79867.96 (± 11.20) 15550.12 (± 0.30) 7003.51 (± 4.38) 0.775 1.037 -1.57

Table 3.
Multiplicity Analysis of 1000-sec Measurements of MOX Standards

The alpha values for MOX standards listed in Tables 2 and 3 are similar to those for pure PuO2. For both the known alpha and the multiplicity analyses, multiplication ranges from 4-8%. Both techniques give comparable accuracies, demonstrating that either technique can be successfully used for PuO2 or MOX if the chemical composition and isotopics are well known, and the precision for the triples rate is good.

To test the effect of poor counting statistics, we also made measurements for 10-sec and 100-sec. For standard 1, the declared-measured values ranged from -25% to +38% for 10-sec measurements and 14% for the single 100-sec measurement using the multiplicity technique - primarily due to the inaccuracy in determining the triples rate. For the known alpha technique, the measured values were consistently within 5% of the declared mass for both 10-sec and 100-sec measurements, demonstrating that the known alpha technique is preferred for samples of known chemical composition if the counting statistics are poor.

We also measured nine MOX scrap samples. The scrap results are summarized in Tables 4 and 5 for the known alpha and multiplicity technique, respectively.

MOX Scrap Sample Assay time (# of cycles * sec/cycle) Totals Rate Reals Rate Multiplication Calculated Alpha Declared- Measured Mass (%)
1 10*60 732337.01 (± 53.6) 42895.47 (± 273.1) 1.000 1.125 -177.12
2 10*60 41929.50 (± 12.9) 6337.26 (± 23.7) 1.017 1.115 -8.39
3 10*60 24625.60 (± 8.8) 5300.21 (± 18.6) 1.082 0.953 52.80
4 10*60 6681.34 (± 3.3) 1007.34 (± 3.6) 1.002 0.953 15.02
5 10*60 154906.80 (± 15.2) 12110.57 (± 92.8) 1.000 0.994 -11.21
6 10*60 166580.73 (± 19.2) 22737.46 (± 102.2) 1.000 1.027 -27.01
7 10*60 745320.28 (± 31.3) 203266.67 (± 596.5) 1.153 1.019 0.01
8 10*60 1129615.08 (± 49.8) 147793.92 (± 551.9) 1.000 1.066 -174.85
9 10*60 74901.54 (± 14.6) 19229.96 (± 71.5) 1.097 0.661 13.84

 Table 4.
Results of known alpha analysis for MOX scrap material.

MOX Scrap Sample Singles Rate Doubles Rate Triples Rate Multiplication Calculated Alpha Declared- Measured Mass (%)
1 737181.73 (± 53.7) 44110.50 (± 274.7) 30488.8 (± 3236.8) 1.07 27.428 54.5
2 41975.03 (± 12.9) 6364.05 (± 24.0) 2641.48 (± 62.6) 1.026 1.229 -1.72
3 24641.73 (± 8.8) 5316.15 (± 18.7) 2799.73 (± 39.0) 1.057 0.745 46.08
4 6682.60 (± 3.3) 1008.63 (± 3.6) 413.09 (± 5.5) 1.024 1.222 27.20
5 155452.79 (± 15.3) 12263.65 (± 94.1) 5229.93 (± 288.7) 1.024 3.762 14.83
6 167203.64 (± 19.2) 23041.39 (± 105.6) 11749.66 (± 558.5) 1.049 1.916 11.09
7 750184.31 (± 31.3) 208886.08 (± 612.7) 171860.11 (± 2786.7 1.133 0.814 -13.64
8 1131135.73 (± 48.8) 150568.91 (± 552.5) 126914.87 (± 11692.4) 1.122 6.138 23.55
9 75041.30 (± 14.7) 19367.05 (± 72.5) 11142.81 (± 197.0) 1.071 0.496 2.23

Table 5.
Results for multiplicity analysis for MOX scrap material.

For the known alpha technique (Table 4), the differences between the declared and measured masses are significantly worse than those for the MOX standards (Table 5). This is as expected because the chemical composition and the declared masses of the scrap are not as well known as those for the standards. For most of the verification measurements for MOX scrap material listed in Tables 4, the calculated alpha values were larger than expected for pure MOX, indicating a third unknown. Therefore, the known alpha technique is not valid for these samples. For samples 1, 5, 6, and 8 of the known alpha analyses, the multiplication values are 1.000. The MULTI software sets multiplication to 1.000 if the calculated value is <1, indicating that alpha is larger than expected - further proof that the known alpha technique is not valid.

For the multiplicity technique (Table 5), the differences are smaller than the known alpha technique. In particular, samples 1 and 8 which had the largest difference using the known alpha technique are significantly improved with the multiplicity technique. MOX scrap sample 1 gave an alpha value of 27.43 but the error in the reported mass was very large, indicating a questionable verification result. We measured the isotopics of the sample using Multi-Group Analysis Code (MGA) to confirm the declared values to minimize the error from the isotopics. The measured results showed good agreement with the declared values. The high singles rate compared to the doubles rate for sample 1 in Table 5 indicates that the sample could contain fluoride. Other possible explanations include the presence of unknown alpha emitters or an unexpected chemical form for the plutonium. MOX scrap sample 1 was re-assayed for 12 runs of 600 sec each to improve the precision of the triples rate. The new assay gave an alpha of 15.923. The difference between the declared and measured mass was 22.83%. This test suggests that samples with large alpha values that give poor results in the PSMC in 600 sec count times can be re-assayed for longer count times to improve the accuracy of the multiplicity analysis. The count time per run should also be increased to improve the precision of the triples rate. This measurement also demonstrated that the multiplicity electronics and analysis can provide useful results for alpha values in excess of 10. There was not sufficient time to repeat all of the other scrap samples at longer count times.

IV. Conclusion

The test results lead to the following conclusions.

  • If the value of alpha is known (e.g. the isotopics and chemistry are well known) then it is possible to perform a correction for the multiplication effects with only two measured values.
  • If there are three unknowns, fissile mass, multiplication, and alpha, multiplicity analysis is required. When the (?,n) emission rate is uncertain due to poorly known isotopics or uncertain chemistry the standard known alpha technique fails. By adding the third measured value it is possible to determine the (?,n) emission rate. With the measured value for (?,n) neutron emission, the plutonium mass can be measured without complete knowledge of the chemical form, but there are restrictions on the count rates. Assay of MOX scrap with alpha values as large as 15 were demonstrated using the PSMC. For higher ratios the (?,n) counting statistics become worse and the accuracy of the measurement suffer.