Presented at the HPS Mid-Year meeting in Alberquerque January 1999 |
Markku Koskelo* , Andrei Savlov**, Dmitry Vzorov*** and Ray Gunnink****
*CANBERRA Industries, Meriden, CT, USA
** Institute of Physics and Power Engineering, Obninsk, Russia
***CANBERRA-Packard, Moscow, Russia
****Consultant, Fremont, CA, USA
Introduction
Fast and accurate determination of the uranium enrichment and/or the isotopic composition of
plutonium is essential in nuclear disarmament and stockpile stewardship. This includes the
traditional verification and confirmatory measurements of Special Nuclear Materials (SNM) like
those performed by Euratom and the IAEA, as well as Material Protection, Control and
Accountability (MPC&A) applications for SNM identification at the nuclear facilities by their
operators. High resolution gamma spectrometry coupled with the Multi Group Analysis (MGA)
code (Gunnink 1987, 1990) has been used successfully for this purpose by both Euratom and the
IAEA for years. The performance of its various versions has been tested extensively (D.
D’Adamo et al. 1990, S. Abousahl et al. 1996). The results are generally quite accurate
regardless of sample age, geometry, chemical composition or attenuation between the sample
and detector.
However, older versions of MGA often required skilled and highly qualified personnel to
correctly set up the instrument and to interpret the results and messages. Together with Dr.
Gunnink, CANBERRA has enhanced the MGA code to be more robust for situations where the counting statistics are far from ideal (Verplancke et al. 1995). In addition, the hardware typically
used for field plutonium measurements has been improved to provide better resolution and
stability (Koskelo at al. 1997).
Recently, materials have been encountered that are stored in lead lined containers that greatly
attenuate all low energy radiations. Although previous versions of MGA allowed the user to
include data from the high-energy region (200 - 1020 keV) in the analysis process, they required
the use of two separate detectors since the low energy region was still the primary region of
analysis. Therefore, a new analysis mode has recently been added to MGA to allow a complete
analysis to be made based on only the high-energy region of a spectrum. In addition to this highenergy
region only mode, the newest version of MGA can still analyze the plutonium
abundances in the traditional single low energy spectrum, and the combination of a low energy
and a high energy spectrum modes. But it is the new high-energy only analysis mode that allows
for the measurement and analysis of highly attenuated samples. In this paper, we will compare
the results from low energy only and high-energy only measurements and show that good results
can be obtained with this new analysis mode.
Experimental
The facilities at Institute of Physics and Power Engineering (IPPE) have different plutonium
samples for test purposes. These include plutonium metal (with 239Pu content of 88% and 95% )
and some special samples of PuO2 with 239Pu content of about 67%, 77% and 87%. The
plutonium metal samples are in the form of disks in stainless steel cladding of 47 mm diameter
and 3.16 mm in height. The plutonium oxide sample is contained in a 6 mm diameter tube and covered by stainless steel foil. The main characteristics of the samples used for the
measurements in this study are shown in Table 1.

From each sample type we chose an item for measurement and analyzed it for its isotopic
composition and compared the results with the known, so called “passport” values. The
“passport” data of the plutonium samples is shown in Table 2. The declaration dates for the
samples 981, 1124 and P37 are June, 1973, July, 1969, and February 1, 1988, respectively.

Two types of High Purity Germanium detectors were used, a planar detector, CANBERRA model
GL0510R with a resolution (FWHM) at 122 keV of 550 eV, and a coaxial detector, CANBERRA
model GC1818 with resolution of 680 eV at 122 keV and 1.73 keV at 1332 keV. Low energy
spectra were collected with the planar detector and the high-energy spectra with the coaxial
detector. A CANBERRA Inspector MCA was used to collect the data and provide the signal
processing for both detectors. For the “classic” low energy analysis with MGA we used a gain
setting of 0.075 keV/channel, and a shaping time of 2 µs. The dead time was 15-20% depending
on the sample and the presence of lead shielding. For the high-energy only analysis we used a
gain setting of 0.250 keV/channel, and kept the shaping time at 2 µs. The dead time for these
measurements was 15-35%.
Each sample was measured with and without a thin lead shield between the sample and the
detector. For the unshielded measurements we used two or three 0.3 mm Sn plates during the
measurements of the plutonium disks to reduce the effect of the 59 keV 241Am gamma radiation.
For the shielded measurements, the thickness of lead plates was selected to reduce the intensity
of all gamma rays below 210 keV. For samples 981 and 1124, we used 3 mm of lead, for sample
P37 we used 4 mm of lead. Three series of experiments were performed for each sample for both
the shielded and unshielded measurements: spectra with poor statistics, with average statistics,
and with good statistics. This was achieved by varying the live time from sample to sample from
a few minutes to about 60 minutes.
Results and Discussion
The measurement results for the non-shielded samples for both the “classic” low energy MGA
analysis and the new high-energy MGA for test sample 981 are shown in Table 3.

The top part of the table shows the results for measurements with a short count times and the
bottom part the results for measurements with long count times. The known “passport” values
are shown in the second column for both portions of the table. The column marked “Classic”
MGA shows the results for the normal low energy MGA analysis followed by the percent
difference relative to the “passport” values. The last two columns show the results obtained with
the new high-energy MGA mode and their percent difference from the “passport” values. Note
that we have purposely excluded the 242Pu results from these tables. It cannot be measured
directly, and a comparison of its reported values to the “passport” values is not a measure of how well MGA can analyze the spectrum but rather a measure of how well the correlation equation
happens to match these particular test samples. In addition, each MGA result column represents
results from a single measurement from the multiple measurements that were made for each
sample.
Overall, the results are quite good for the main isotopes and there is not much difference between
the poor statistics and the good statistics. The uncertainties reported by the two modes of MGA
are not shown in the table. However, a statistical analysis indicates that they are appropriate for
the observed deviations from the “passport” values. The results for sample 1124 show very little
difference the poor statistics and the good statistics, and have not been included here. The results
for test sample P37 are shown in Table 4. For these results, the difference is a little bit more
noticeable, particularly in the high-energy analysis.

The “classic” MGA analysis cannot be performed on the measurements with the lead-shielded
containers. All other techniques that use low-energy regions have the same problem. The lead lining
prevents the low energy gamma rays that they use as a basis of the analysis from reaching the
detector. However, the new high-energy only MGA analysis has no problems with the lead lined
samples as long as the gamma rays above 300 keV can be detected and measured. The results with
the high-energy analysis mode for test samples 981, 1124 and P37 encased in lead are shown in
Tables 5, 6 and 7, respectively. Note that each of the MGA results columns again represents only
one measurement of the multiple measurements that were made with each sample.



It can easily be seen from the tables that there is generally an improvement in the analysis results as
a function of the statistics. But it should be noted that even the good statistics spectra were all
measured for less than 30 minutes. Again, the reported uncertainties from the MGA analysis are not
shown on these tables for the sake of clarity. However, just like in the case of the unshielded sample
measurements, a statistical evaluation of the replicate measurement results and a comparison of the
reported uncertainty against the deviation from the “passport” values shows that the uncertainties
reported by MGA are appropriate.
Note that the declared date for two of the samples is 25 to 30 years ago. This means that the
241Pu has gone through nearly two half-lives of decay. The comparisons made here were done
simply by decaying the measured results back to the declared date for convenience, since the
MGA code easily accommodates such a calculation. However, with such old material, most of
the 241Pu has decayed to 241Am. This presents a problem in the 600 keV region which is crucial
for the measurement of 238Pu and 240Pu in the high-energy mode. The problem is that the 662
keV peak of 241Am, and it's Compton continuum, begin to mask the region that must be analyzed. However, this is a problem that applies to all codes using the high-energy region - not
just to MGA.
Conclusions
The measurement results for the unshielded samples show that both modes of MGA are capable
of analyzing the data effectively. Due to the spectral characteristics, the high-energy analysis
generally exhibits somewhat larger deviations from the “passport” values than the low energy
analysis for the same amount of count time. At the same time these larger deviations are
accompanied by larger error bars for the reported values. A statistical evaluation of the
uncertainties of replicate measurements shows that the uncertainties reported by MGA in either
mode are appropriate relative to the deviations from the “passport” values. It is well known that
the lead-shielded samples cannot be analyzed with any of the “classic” low energy analysis
methods, including older versions of MGA. However, with the new high-energy mode capability
that has been incorporated into the latest version of MGA described here, this limitation has been
removed.
References
R. Gunnink, “A New One-Detector Analysis Method for Rapid High-Precision Plutonium
Isotopic Measurements”, Proceedings of the 9th ESARDA Symposium on Safeguards and
Nuclear Material Management, London, UK, 12-14 May 1987: 167-171.
D. D’Adamo, M. Franklin, S. Guardini, C. Vicini, R. Gunnink, W. Ruhter, and G. Varasano,
"Performance Evaluation of Gamma Spectrometry Codes for International Safeguards",
Proceedings of INMM, Vol XIX, 1990: 782-791.
S. Abousahl, A. Michiels, M. Bickel, R. Gunnink, and J. Verplancke, Nucl. Instr. & Meth. in
Phys. Res. A368 (1996) 449-456.
J. Verplancke, D. Davidson, M. Koskelo, R. Gunnink, J. L. Ma, J. Romeyer-Dherbey, S.
Abousahl and M. Bickel, "Applying MGA for Waste Characterization". Proceedings of WM95,
Feb 26 - Mar 2, 1995, Tucson, Arizona, Session 17, Paper 7.
M. J. Koskelo and G. Gardner, “An Improved U-Pu System for Field Pu Measurements”,
Proceedings of the 19th ESARDA Annual Symposium on Safeguards and Nuclear Material
Management, Montpellier, France, May 13-15, 1997: 441-445.