Presented at WM ’05 Conference February 27 – March 3, 2005 Tucson, AZ |
Frazier L. Bronson CHP
CANBERRA Industries, Inc.
800 Research Parkway, Meriden, CT 06450 USA
Valery Atrashkevich Ph.D.
Verdansky Institute of Geochemistry and Analytical Chemistry
Moscow, Russia
INTRODUCTION
Large NaI detectors are frequently used for high
efficiency measurements of low levels of gamma
emitting radioactivity. The most common sizes of
these detectors are 4”x4”x16” and 3”x5”x16”. The 4x4
detector was the initial replacement for large multitube
cylindrical detectors, and is commonly used for
geological surveys, while the 3x5 detector is normally
used in CANBERRA systems, as it is similar in cost and
background, but approximately 25% higher in efficiency.
At CANBERRA, our most common use of these
large rectangular detectors is in our Whole Body
Counting systems. The FASTSCAN counter uses two
of them, the ACCUSCAN scanning bed counter can
use up to four, and we have also built special in vivo
counters using up to 32 of them. Animal counters also
used these large detectors as well as counting systems
for waste in drums, boxes, and trucks. These
detectors are also commonly used to survey large
volumes of soil or building debris in D&D and ER
projects. This is commonly done in situ with fixed or
moving detectors, on conveyors with the sample
moving passed a fixed detector, or with the sample in
large containers e.g. trucks. There is also much
interest in use of arrays of these large NaI detectors
for Homeland Security portal monitors for pedestrians
and vehicles.
Large detectors are more sensitive than smaller detectors,
and better able to detect small levels of
radioactivity. Traditionally large area gamma detectors
have been plastic scintillators. The advantage of
NaI as compared to plastic scintillators is that gamma
spectroscopy can be performed to identify and to
quantify the radionuclides from the item or area being
measured. But, before this can be done, the system
has to be calibrated for efficiency as a function of
gamma energy.
The normal way to perform efficiency calibrations is to
construct a calibration source that is the same physical
size, and constructed in a radiologically identical
manner to properly simulate the item being measured.
Then, a known amount of radioactivity is
distributed in the same manner as it will be in the item
to be measured. This must be done for a wide range
of energies to establish a calibration curve. For small
water samples in a laboratory this is relatively simple.
But, as the samples get larger, or if they are not liquid,
the task becomes more complicated, more expensive,
more time consuming, and less accurate. For
these situations mathematical efficiency calibrations
are especially attractive.
The code MCNP (Monte Carlo Neutron-Particle) is
widely available for evaluating radiation transport
phenomenon. CANBERRA has previously shown that
MCNP, when properly applied, can create gamma
spectroscopy efficiency calibrations that are accurate
to 5%1. But creating these models takes quite a bit of
time and experience, and running the computations
takes much computer time, especially for large samples
at far distances. Computer time can be hours or even
days.
The CANBERRA ISOCS (In Situ Object Calibration
Software) mathematical efficiency calibration software
was developed to simplify and speed up this
efficiency calibration process2. It is capable of calibration
accuracy in the 5-10% range3. It was developed
originally for Ge detectors but has been extended to
NaI detectors. A critical assumption in this calibration
software is that the detector is a right circular cylinder,
and therefore that the radiation response is radially
symmetric about the detector axis. This, of course, is
not true for these rectangular detectors. But perhaps
the ISOCS calibration technique is still good enough
to be useful for calibrations of adequate quality in a
limited spatial region around the detector.
This investigation was designed to answer that question.
We know from extensive testing that the ISOCS
software can produce results that are within 2%
agreement of the MCNP results, for the specific
shapes that are allowed within the various ISOCS
templates. Therefore, if this investigation done using
MCNP shows acceptable agreement, then so should
the ISOCS process.
What does “good agreement” mean? A value of ±20%
was subjectively chosen. Gamma spectroscopy using
NaI detectors is not nearly so easy as with high
resolution Ge detectors. Because of the much poorer
peak resolution of NaI, errors in determining the net
peak area from other peaks in the spectrum can easily
occur. While laboratory users measuring simple spectra
can get results better than 20%, it is difficult with
multiple nuclides in the spectra (e.g. background
containing radium, thorium, potassium), and also at
low levels as typically are encountered for the applications
with these large detectors.
METHOD USED
The first step was to create an MCNP model of the
reference 3”x5”x16” NaI detector, and then a model of
an “equivalent” cylindrical detector. The rectangular
detector was modeled using the known dimensions
and materials of the real detector. The cylindrical
detector was modeled using the same exterior protection
materials outside of the NaI, however the diameter
and thickness of the NaI were optimized to best
match the efficiency of the rectangular detector for a
thin source that is 2 m diameter and 1 m from the front
face of the detector. This matching was done by
iteratively varying the diameter and thickness, each time evaluating the agreement between the cylindrical
and rectangular detector. The energies used were
100, 500, and 2000 keV. When the process was
stopped, the agreement was within 1% at all energies
for a cylinder that is 26.3 cm diameter and 6.9 cm
thick. Figure 1 shows the geometry and the results
graphically.
POINT SOURCES
The next experiment was to use MCNP to compute
the efficiency for both the equivalent cylinder and the
3”x5”x16” rectangular NaI at 50 different point locations
in front of the detector. These were all placed in
radial symmetry in one quadrant about the axis of the
cylindrical detector. Points were placed in planes
located at 1 cm, 10 cm, 1 m and 10 m from the front
face of the detector. They were placed at 0 degrees
(parallel to the 16” dimension], 17°, 34°, 54°, 72°, and
90°. They extended radially 10 meters for the 1 cm
and 10 cm distances, and 100 meters for the 1 m and
10 m distances. There were 42 points on each of the
six radials; at each of these points the efficiency was
computed for both detectors at 100 keV, 500 keV, and
2000 keV. An analysis of the data showed that as long
as these small sources are greater than 10 cm away
from the face of the detector, and within a 90 degree
subtended solid angle radiating out from a 40 cm
diameter plane on the face of the detector (this
encompasses all of the 3”x5”x16” detector), then the
cylindrical model was within 20% of the rectangular
efficiency for nearly every location. Since that is
where a user would typically place a source, this was
quite encouraging. If the source is outside this area,
the worst case error is a factor of 2. Figure 2 shows a
selected portion of this data graphically. The best
agreement was at 17° and the worst is at 0° and 90°.
 |
Figure 1.
The optimization geometry for developing the equivalent-sized cylinder
that best matches the efficiency of the 3”x5”x16” NaI
detector.
The graph at the right shows the ratio of cylindrical to rectangular efficiency. |
 |
Figure 2.
The point source test geometry and results. |
CIRCULAR AND LINEAR SOURCES
Further computations were done to show calibration
accuracy for non-point sources. A series of circular
planes was the next test. This geometry is a common
one for soil surveys. The planes were at 1 m from the
detector face, and the diameter was varied from 1 m
to 30 m. For cylindrical sources at 1 m, the accuracy of calibrations with the equivalent cylindrical detector
is always within 20% for all source diameters. Based
upon the point source response, it is expected that this
relationship also holds true for circular planes at
further distances, but this was not tested. Figure 3,
upper section, shows this geometry and the data
graphically.
The next test was for linear sources. Linear sources
might be representative of a person standing, a pipe,
sources moving on a conveyor, or a moving car or
train. A distance of 1 m from the face of the detector
was used, and the line length was varied from 1 m to
30 m. The line was positioned at 0° and 90°. These are
the two worst case situations, all intermediate angles
will have better agreement. The accuracy is within 20% for all three energies, as long as the source
length is less than 3 m. In all cases the error was worse
at 100 keV than the other two energies. Based upon
the point source data, it is speculated, but not tested,
that as the line source is moved further away from the
detector, the length that is within 20% accuracy will
increase approximately proportionally. Figure 3, lower
section, shows this geometry and data graphically.
 |
Figure 3.
Geometry and results for circular planar source (upper portion) and for
linear source (lower portion). |
WHOLE BODY COUNTER
The last test was to simulate the CANBERRA FASTSCAN
Whole Body Counter. This counter is used in most of
the US Nuclear Power Plants, and many other places.
This counter has two 3”x5”x16” NaI detectors in a
vertical linear array. The person stands next to the
detectors, with his back about 40 cm away from the
face of the detector. For expediency, a simplistic
model of a person was created with MCNP. That
model had two cylinders stacked, where the masses
of the cylinders in approximate standard-man proportions.
The radioactivity was distributed uniformly,
simulating a systemic uptake, e.g. 137Cs.
The detectors were also positioned in the model at
approximately the same locations as in the instrument.
Since the spectra are summed, the efficiencies
from the two detectors were summed here. The
agreement was very good – within 5% for all three
energies, as shown in Figure 4.
 |
Figure 4.
Simulation of the FASTSCAN Whole Body Counter. |
CONCLUSION
A series of tests has been done that show that a
properly proportioned cylinder can be shown to have
an efficiency approximately equal to a commonly
used rectangular NaI detector. This approximation
holds true for all small sources located further than
10 cm from the detector, and within ±45 degrees of a
line perpendicular from the face of the detector. This
equivalency holds true for all circular planar sources
out to at least 30 m in diameter. This equivalency
holds true for all line sources up to 3 m in length. This
equivalency is true for a simplistic representation of a
person in a common Whole Body Counter. Sources
within these spatial parameters can be calibrated to
within 20% of the correct value. Sources outside
these spatial regions can be worse, but not more than
a factor of 2 different than the correct efficiency.
Since we know that the ISOCS method accurately
reproduces the MCNP results for the 20 different
shapes that are defined as templates in the software,
we know that an ISOCS model using the same
cylindrical equivalent detector as used here will give
similar results. The ISOCS method allows efficiency
calibrations within seconds, instead of the very long
MCNP computations times (hours-days). Consequently,
this method should be practical for design of
proposed systems, and reasonably accurate calibrations
of large sources, such as waste containers,
in situ soil measurements, moving cars and trucks,
soil on conveyors, and even people.
REFERENCES
1. F. L. BRONSON AND L. WANG, “Validation of
the MCNP Monte Carlo Code for Germanium
Detector Gamma Efficiency Calibrations”, Proceedings
of Waste Management ’96, Tucson, AZ,
Feb. 1996.
2. F. L. BRONSON, “ISOCS, an In Situ System and
Portable Gamma Spectroscopy Lab that can be
Taken to the Accident Site”, Proceedings of the
ANS 7th Topical Meeting on Emergency
Preparedness and Planning, Sept. 1999, Santa
Fe, NM.
3. R. VENKATARAMAN, F. BRONSON, V.
ATRASHKEVICH, B.M YOUNG, and M. FIELD, “Validation of In Situ Object Calibration System
(ISOCS) Mathematical Efficiency Calibration
Software”, Nuclear Instruments and Methods in
Physics Research (A), 422 (1999), pp 450-454.