Presented at the 46th Annual meeting of the INMM, Phoenix, AZ, July 10-14, 2005 |
Richard Thomason and Stephen Croft
Canberra Industries, Meriden, CT, 06450, 803/652-8290
ABSTRACT
Nondestructive assay instruments such as segmented gamma scanners and tomographic
gamma scanners generally use measured transmission values to calculate correction
factors for attenuation effects. Current practice is to use one or more transmission
sources with peak energies close to the assay energy and to use interpolation
techniques to obtain transmissions at each assay energy. This practice limits
the source radioisotopes that are useful for each assay radioisotope and can
lead to using isotopes that are expensive or have half-lives so short as to
require frequent replacement. Use of the Material Basis Set (MBS) approach
to perform the transmission interpolation or extrapolation provides more flexibility
in choice of transmission sources and provides transmission values over a wider
range of assay energies. A previous computational study by the authors [1]
quantified assay errors introduced by the MBS method for four radioactive sources
and eight diverse matrix types assuming exact values of measured transmissions.
The present study includes the effect of statistical uncertainties in measured
transmissions. To determine the effect of these uncertainties on the MBS-determined
values, measured transmission uncertainties were corrected for intensity and
attenuation to estimate measured transmission uncertainties for the different
assay scenarios. Using these estimated uncertainties, the assay bias and random
error introduced by the MBS method plus transmission measurement uncertainties
was determined. Particular attention was given to extrapolation to low transmission
values at low gamma ray energy (60 keV), where the energy variation is the
steepest.
INTRODUCTION
The measurement of the quantity of one or more radioisotopes in a container
by detection of the emitted gamma rays generally requires that a correction
be made for attenuation of the gamma rays by the material containing the radioisotopes.
The correction can be calculated directly if the composition, density, and
uniformity of the material are known along with the activity distribution.
Most often this is not the case, and the correction is made based on measured
transmissions of gamma rays through the container. The usual method is to use
a separate “transmission” gamma ray source that emits gamma rays
at one or more energies near the assay energy. Transmissions through the container
are measured at the transmission source energies, and then these values are
interpolated or extrapolated to the assay energy or energies using exponential,
linear, or quadratic functions or polynomials in logarithm-logarithm space.
Sources with energies most appropriate for assay of some important radioisotopes,
such as 235U and 239Pu, have relatively short half lives and require frequent
replacement. Also, a single transmission source used with traditional interpolation
and extrapolation methods is not well suited for attenuation correction over
a wide range of energies, as needed for an assay instrument that quantifies
many radioisotopes with each assay sequence. Use of the MBS method determines
the transmission through a container over a wide range of energies based on
measured transmissions at only a few strong, disparate energies. Also, this
method can be applied using transmission sources having multi-year half-lives
and needing replacement at no less than 7-year intervals. The calculation-based
study presented here quantifies the assay bias and random errors for each assay
scenario due to attenuation-correction using the MBS and provides guidance
for its use.