REGIONAL ACCESS

EXTRANET ACCESS

TAM Series, Tritium Air Monitor

Canberra's tritium monitors consist of two basic units, one of which is portable and one that is used for fixed applications. Both of these are able to accommodate serial communications with a remote processing and display unit. These units have the capacity to draw air from the atmosphere, pass it through a filter to remove the dust, and measure the tritium concentration in DAC units on a digital/analog display. Measurement ranges vary slightly and all Models have interference immunity from other airborne radioactive gasses and vapors such as Argon, Xenon, Iodine, etc.

  • 2.4 liters Flow-though dual sealed ion chamber
  • Removable filter for particulate contamination
  • Ion trap
  • Gamma compensation
  • Automatic Radon substraction
  • Solid state electrometer
  • Front panel electrometer output (0-5 V DC)
  • Weight: 25 lbs (11.4 kg)
  • For glove boxes, NPPs Neutron generators, Cyclotron and Accelerators
Model:
Applications :
Power:
Range :
Volume :
Advantages :
TAM-100D
Fixed
AC only
0.5 to 106 µCi/m3
7-9 l/mn
Heater for fast cleaning
Quicker decontamination
More Rigorous test/calibration
TAM-73D
Portable
AC & Batteries
1 to 106 µCi/m3
3-5 l/mn
Rechargable Battery
backup for 2 hours

Products in this Category:

QUESTIONS?

In the United States
(800) 243-3955

Outside United States:
(203) 238-2351

Privacy | Quality | Terms & Conditions | Webmaster