Lin/Log Ratemeter - Model 1481LA
Features |
|
|
Description |
The Model 1481LA LIN/LOG Ratemeter NIM module offers the user a choice of either linear or logarithmic presentation of the average number of pulses per second being received. In the linear mode, the Model 1481LA features ten full-scale ranges, from 10 to 500 x 103counts per second, permitting accurate reading at any range, or the choice of an extremely wide dynamic input range. In this mode, the input pulse rate may be integrated over any one of four switch-selectable time constants, from 0.5 to 40 seconds. In the logarithmic mode, the Model 1481LA features two three-decade ranges, from 10 to 104counts per second, and from 102 to 105 counts per second, allowing a wide dynamic input range to be monitored without changing the control settings. The Model 1481LA’s alarm relay is energized whenever the count rate exceeds a pre-selected linear percentage of any full scale range. The Model 1481LA offers ±0.5% linearity, a rear panel 0-100 mV recorder output, and a 3% precision analog front panel meter. |
Note to our International Customers: If you would like a quote on this product, please contact your Local Sales Office for assistance. |
||
| Models | Description | |
| 1481LA | Lin/Log Ratemeter | |
QUESTIONS? 
In the United States
(800) 243-3955
Outside United States:
(203) 238-2351