Detector LN2 Monitor
For more detailed information click here  > >
Product Spec Sheet PDF Format

Model 1786A Detector LN2 Monitor - Click for Larger ImageFeatures

 
  • Automatic protection against accidental detector warmup
  • Delayed shutdown saves experiments
  • Audible, visual and contact closure outputs

Description

The Model 1786A Detector LN2 Monitor protects LN2-cooled detector systems against accidental warmup by monitoring liquid level in the Dewar and providing audible and visual warning of low liquid levels. It provides a signal for remote shutdown of Detector Bias Supply which has an Inhibit input. This singlewidth NIM module is suitable for most dipstick and integral type cryostats, but is not suitable for portable cryostats.

When your LN2 sensor detects a problem, the LN2 Monitor has provisions for turning off high voltage when an alarm condition occurs to protect detectors and electronics from damage. Shutdown does not occur until at least 20 minutes after the condition is sensed, and well before total depletion of the liquid nitrogen, so ongoing experiments can continue without interruption.

The Model 1786A must be manually reset following a power interruption. This will prevent HV Bias from coming on automatically (and abruptly) when the power is restored.


Note to our International Customers: If you would like a quote on this product, please contact your Local Sales Office for assistance.

Models Description
1786A Detector LN2 Monitor