Proportional Counter Preamplifier - Model 2006
Features
- Low noise design: <350 ion pairs at Cs = 0 pF
- High charge rate capacity for high count rate applications; up to 2 x 10-7 coulombs/second
- FET input, diode protected
- Fast risetime: <20 ns at Cs = 0 pF
- High voltage operation up to ±2000 V dc
Description
The CANBERRA Models 2006 preamplfier is designed for use
with proportional counter detectors. The preamp converts the
ionization charge from the detector into a step voltage pulse
whose amplitude is proportional to the total charge collected
in each event. The pulse decays with a time constant of 50
ms for presentation to a main
amplifier.
For typical use with a positive HV bias to the anode of the detector, the preamplifier generates positive polarity energy pulses which are shaped in the amplifier as necessary for counting or low energy x-ray analysis.
This preamplifier contains a diode-protected FET input stage
optimized for low noise operation. The first stage acts as an
operational integrator which produces an output potential
proportional to the accumulated charge. This integrator
configuration is followed by a pole/zero adjustment to return
the unipolar pulse to a reference or baseline level without
overshoot, and a differentiator to provide the 50 ms tail pulse. The second active stage operates
as a buffer for driving various cable lengths without pulse
degradation. Conversion gains of nominally 47 mV/M-ion-pairs or
235 mV/M-ion-pairs may be selected by a jumper plug inside the
unit. The higher scale factor is especially useful for best
signal-to-noise ratio in experiments involving low energy
sources.
The noise level is equivalent to less than 350 ion pairs with zero source capacitance using 2 ms near-Gaussian pulse shaping, and degrades less than 1 ion pair per pF of input capacitance. As illustrated in Table 1, low noise and fast rise time prevail over a wide range of input capacitance. The high count rate capability of the design is evidenced by a charge rate capacity greater than 2 x 10-7 coulombs per second. In order to take full advantage of such a high count rate capability, a high count rate main shaping amplifier is recommended.
A test input is provided to assist in system setup. The test capacitor value is certified in the unit test report for reference use as a secondary charge calibration standard. The nominal voltage gain from the test output is X 1 for the output scale factor of 47 mV/M-ion-pairs and X 5 for the output scale factor of 235 mV/M-ion-pairs.
Power is supplied from the associated CANBERRA pulse shaping amplifier. The power lines are filtered within the unit to provide high noise immunity. A 3 m (10 ft) power cable is provided with the preamp.
Note to our International Customers: If you would like a quote on this product, please contact your Local Sales Office for assistance. |
||
| Models | Description | |
| 2006 | Proportional Counter Preamplifier | |