JCC-31: High Level Neutron Coincidence Counter


View full specifications (PDF)

jcc_31.jpg 
(5257 bytes)Features

  • Designed for passive neutron coincidence assay of plutonium
  • Eighteen 3He detectors
  • Fast Amptek electronics
  • Transportable for inspections at multiple sites
  • Optional transport container

Description

The JCC-31, which is based on a technology transfer from Los Alamos National Laboratory, measures the 240Pu-effective mass in a sample by detecting coincidence neutrons from the spontaneous fission of plutonium. The effective mass of 240Pu is the mass of 240Pu which would emit the same number of spontaneous fission neutrons per second as the combined 238Pu, 240Pu and 242Pu in the sample. The detector can measure up to several kilograms of plutonium. A JSR-12 Neutron Coincidence Analyzer, a computer and analysis software are required for coincidence counting but are not included with the JCC-31.

The JCC-31 has a cylindrical-shaped sample cavity 41 cm high by 17 cm in diameter. It is intended to assay plutonium samples including PuO2, mixed oxides (PuO2-UO2), metal carbides, fuel rods, fast critical assemblies, solution, scrap, and waste. A cadmium sleeve surrounds the sample cavity to prevent the re-entry of thermalized neutrons into the sample, which could induce fission in the sample and adversely affect the results. Outside the cadmium sleeve is a ring of high-density polyethylene with eighteen 3He tubes placed in the polyethylene.

The tubes are arranged in a single ring around the sample with optimum spacing between the tubes for maximum counter efficiency for a transportable counter. The tubes are divided into six groups of three with each group wired together and connected to one JAB-01 Amplifier/Discriminator circuit board. The six JAB-01s are mounted inside a sealed junction box. LED indicator lights are placed externally on the junction box to indicate proper operation of each JAB-01 channel. Electrical connections between the JCC-31 and the JSR-12 include +5 V, HV, and a single “ORed” output signal.

A cadmium sleeve wrapped around the outside of the JCC-31 provides radiation protection for personnel as well as background reduction.


View full specifications (PDF)

Top of Page

Note to our International Customers: If you would like a quote on this product, please contact your Local Sales Office for assistance.

Models Description
JCC-31 High Level Neutron Coincidence Counter


Product Inquiry / Request

In the United States
(800) 243-3955

Outside United States:
(203) 238-2351