FRAM; Isotopics Software


View full specifications (PDF)

Features

  • Determines relative plutonium isotopic abundances or uranium isotopic abundances in non-destructive assay applications
  • Uses spectra collected with HPGe detectors
  • Incorporates a sophisticated peak fitting and multiplet deconvolution algorithm to analyze complex peak regions
  • Requires no efficiency calibration for matrix density, thickness or container characteristics
  • Developed by Los Alamos and licensed to CANBERRA

Description

FRAM (Fixed-Energy, Response Function Analysis with Multiple Efficiency) is a code developed by Los Alamos to analyze pulse-height spectra generated by high-resolution gamma-ray detectors. It has been used primarily to determine the isotopic composition of plutonium in special nuclear materials. Its design and flexibility allow it to easily measure ratios and distributions of isotopes other than plutonium in arbitrary items.

In nuclear safeguards, FRAM complements calorimetry and neutron coincidence counting by allowing measurements to be interpreted in terms of total plutonium mass.

Table 1. Material Types Measured with FRAM

240Pu (2% – 95%) of total plutonium

241Pu – 237U (nonequilibrium)

Am/Pu in heterogeneous combinations

241Am (0.01% to >50%)

235U/Pu mixed (MOX) ratios from 0.005 – 35

243Am – 239Np

244Cm

237Np

234U, 235U, 238U relative abundances, (no plutonium present), 0.2 to >97% 235U

238Pu (80%)

Materials in shielded or heavy-walled containers
(³13 mm Pb for plutonium, ³16 mm steel for uranium)


Measurement Principles

In its principal application, FRAM analyzes photopeaks in the spectrum of plutonium gamma rays detected by a high-purity germanium detector. This spectrum contains peaks from the plutonium isotopes 238-241Pu, 24lAm and often other isotopes such as 235U or 237Np. FRAM combines this information to produce isotopic ratios independent of sample size, shape, physical and chemical composition, measurement geometry, and container characteristics. The results are obtained using only the spectral data and known fundamental nuclear branching fractions and half lives and do not require calibration with standards.

A requirement that the plutonium in the sample has one isotopic composition is usually satisfied. However, certain important cases in which the 241Am is not homogenous with the plutonium can also be analyzed using the “multiple efficiency” characteristics built into the FRAM code. This allows FRAM to improve the results on elementally (Am/Pu) heterogeneous pyrochemical residues. The multiple efficiency feature of FRAM can also be applied to other types of elementally heterogeneous materials. The physical and chemical characteristics of the measured item need not be uniform or even well known. Items may contain mixtures of solids and powders with no ill effect as long as the plutonium is isotopically uniform.

FRAM is primarily an analysis package, although it is capable of simple control of data acquisition hardware as well (see below). The end results are displayed in their own window as shown in Figure 1. It is also possible to display the spectrum in various ways, including showing the fitted results on each peak region.  FRAM also contains an MCA emulator display and displays the spectrum during data acquisition.


Figure 1.
Example of spectrum display and analysis result printout capabilities

Performance

The isotopic fractions themselves are usually the desired result for quality control and process measurements. For nuclear safeguards, two additional quantities are most often the end results of the isotopic measurement process. These integral quantities, the effective specific power Peff(W/g Pu) and the effective 240Pu fraction 240Pueff, are derived from the measured isotopic fractions, and known nuclear constants.  They are used to interpret calorimetry measurements (Peff) and neutron coincidence counter measurements (240Pueff) respectively.

Over a wide range of measurement and sample conditions FRAM performance falls into ranges shown in Table 2.

Table 2. Typical Ranges of FRAM Measurement Characteristics

Measurement Time

20 min to 1 hr

Sample Size

100 mg to many kg
(limited by criticality)

Precision
(relative standard deviation)

Peff – 0.2% to 0.5%

240Pueff – 0.8% to 3.0%

Bias

Peff – <0.3%

240Pueff – <1.0%

Individual isotopes – <1.0%
(238Pu to 241Pu)

MEASUREMENT HARDWARE

While FRAM is primarily an analysis package, it does include a simple method for data acquisition. It can set the preset, start and stop spectrum acquisition, and save the spectra to a file using the following CANBERRA MCA devices:

DSA-2000 Digital Spectrum Analyzer
Model 556A AIM with ICB NIM
InSpector Portable MCA
InSpector 2000 DSP Portable MCA
AccuSpec Boards
System 100 Boards

The details of the MCA setup are defined with the Genie-2000 MID Editor Setup Wizard. This utility program leads the user through a simple set of selections to define the MCA configuration. More complex cases are defined using the MCA Input Definition Editor itself.

If the front end electronics are not programmable, the use of the MID Editor Setup Wizard, or the MID Editor itself is usually all that is required to define the configuration. The FRAM code can then be used as a stand-alone package. If the front end electronics are programmable, they need to be setup using the Genie-2000 software since FRAM does include any controls for such adjustments.

OPERATING PLATFORMS

FRAM is available to work with the Genie-2000 software suite on the Windows 95, Windows 98 and Windows NT operating systems.