Ratemeters, Probes & Accessories

Products in category

Radiation monitoring systems are deployed for a diverse set of applications in many types of facilities including nuclear power plants, fuel cycle facilities, research reactors, national laboratories, universities and medical labs. While the activities conducted at these facilities are quite different, the need for reliable, robust, easy to use radiation monitoring systems is shared. CANBERRA SafePoint™ RMS solutions offer modular hardware and electronics that support a broad spectrum of area monitoring, liquid and air monitoring scenarios and are easily scalable from small to large configurations.

Configuration Guide – RMS Detectors and Ratemeters for Area Monitoring
The specific requirements of each application will dictate the type of the SMART detectors, ratemeters, support hardware, alarms units and supervisory software needed. Click here (or the button below) for a brief guide that provides the tools for simple configuration of the radiation monitoring solution that's right for you.
ADM606 Rackmount Multifunction Ratemeter/Scaler

ADM606 Rackmount Multifunction Ratemeter/Scaler

Description

The ADM606 is a microprocessor controlled digital ratemeter, designed to power, control and process data from various types of CANBERRA radiation detectors. The ADM606 can operate with both area and process monitoring detectors for measurement of many types of low and high intensity radioactivity. The area monitoring detectors are designed for open area detection of gamma or neutron radiation. Process monitoring detectors detect alpha/beta/gamma activity and are designed to mount in fluid or gas flow streams such as ventilation ducts, stacks or liquid piping systems.

The ADM606 is designed to take full advantage of the features of the CANBERRA "SMART" detector series, it also can operate with conventional detectors, or with a combination of conventional and "SMART" detectors. When coupled with the "SMART" detectors, the ADM606 interrogates each connected probe to determine the detector type and calibration constants, then automatically sets the display of measurement type and units for that measurement channel. This feature allows for quick detector swapout, and permits factory calibration of detectors rather than time consuming in-field calibration of each monitoring station.

The ADM606 can be installed as one of a total of three units mounted in a 19 in. wide by 7 in. tall CR600 NIM style rack. The ADM is interfaced directly to a Rear Termination Panel Assembly, CR600TP. The CR600TP and optional CR600CIP, Customer Interface Panel, provides terminal blocks connections to relay contacts, communication ports, and analog and digital inputs/outputs.

The ADM606 provides the following user interface features: 2 lines x 20 characters vacuum fluorescent display, six button membrane switch keypad, three position key switch, four LED indicators, and a RS-232 serial port.

The ADM606 also provides the following control and communication circuitry with connections made through the Termination panel, CR600TP: Eight Single Pole Double Throw (SPDT) relays.

The optional Customer Interface Panel, CR600CIP provides terminal block connections for: Five Double Pole Double Throw relays (DPDT) and one Single Pole Double Throw relay, four Isolated 4-20 mA analog outputs and three Isolated RS-485 communication ports. The Analog Outputs are proportional to the dose rate for area monitors or count rate/concentration for process monitors.

The ADM606 allows configuration of the alarms to be latching or non-latching and to configure the alarm relays as normally energized (de-energized on alarm, failsafe) or normally de-energized (energize on alarm).

In latching mode, the alarm indicators continue to indicate alarm until after the rate drops below the alarm setpoint and the operator has pressed the Reset button. In non-latching (or tracking mode), the alarm indicators automatically reset when the rate drops below the alarm setpoint.

When used in the local configuration, ADM606 supports CANBERRA's unique Time-to-Count technology and LED-based gain stabilization for scintillation detectors for enhanced RMS performance.

The ADM606 can be used in a remote configuration in a control room where it can interface with other field-installed ratemeters such as the iR7040, ADM606M, ADM616 or Tritium Monitors such as the TAM100D/T100DSI, or TAM73D/T73DSI.

Three ADM606 Ratemeters shown in CR600 rack configuration.

Features

  • Stand-alone operation or integrated computer-based system
  • Connectivity to both area and process monitor type probes – Alpha, Beta, Gamma and Neutron detection – low and high intensity radiation fields
  • Supports simultaneously up to three "SMART" detector probes
  • Dependable and user friendly interface features, including SI units display
  • Key switch to prevent unauthorized use or changes to settings
  • Extended control and communication, serial RS-485
  • Analog and digital inputs/outputs
  • Self diagnostic and minimum maintenance
  • Data protection with non-volatile memory and lithium battery backup
  • Two operational modes – Local or Remote
  • Custom firmware configurations supported

Quality

  • 10CFR50 App. B QA, non-safety or optional 1E safety
  • Seismically qualified
MDN455EPA/MDNR455EPA - Beta/Gamma Scintillation Detector/Preamplifier Set

MDN455EPA/MDNR455EPA - Beta /Gamma Scintillation Detector/ Preamplifier Set

Description

The MDN455EPA/MDNR455EPA matched pairs consist of a beta/gamma scintillation detector coupled to a PA300E preamplifier. While the MDN455EPA model utilizes a straight cable connection at the back of the detector, the MDNR455EPA utilizes a right angle connection for use in tight spaces. These factory matched and calibrated detector pairs simplify deployment in new or existing systems by eliminating the need for calibration/matching during installation.

The MDN455E beta scintillation detector consists of a plastic beta scintillator and a non-hygroscopic gamma scintillator optically coupled to a photomultiplier (PM) tube via a Lucite light pipe. A reference Light Emitting Diode (LED) and a temperature sensor are imbedded in the Lucite pipe.

This detector operates with a unique gain stabilization circuitry for temperature- compensated drift-free operation, resulting in improved accuracy and extended calibration cycles.

This detector operates in conjunction with a preamplifier, Model PA300E. The preamplifier contains circuitry that compares the LED pulses with a reference to provide closed loop feedback for automatic gain compensation. The PA300E also amplifies and shapes the detector output signal in order to provide digital pulses to the remotely located ratemeter. Procurement of matched detector/preamplifier sets is recommended to simplify field maintenance and system support. See ordering information for details.

The MDN455EPA/MDNR455EPA can be connected to multifunction control and display unit such as iR7040, ADM606, and ADM616.

MDN455EPA/MDNR455EPA - Beta /Gamma Scintillation Detector/ Preamplifier Set

Features

  • Automatic gain stabilization using built in LED
  • Operates in pulse mode and extended range current mode
  • Linear over 8+ decades
  • Calibration cycles of up to two years
  • Detectors are packaged in a water resistant O-ring sealed cylinder with a mu-metal shield
  • Includes calibrated and matched pair PA300E preamplifier with three single channel analyzer channels
MDN55EV1PA/MDNR55EV1PA - High Sensitivity, Extended Range Scintillation Detector/Preamplifier Set

MDN55EV1PA/MDNR55EV1PA High Sensitivity, Extended Range Scintillation Detector/Preamplifier Set

Description

The MDN55EV1PA/MDNR55EV1PA matched pairs consist of a gamma scintillation detector coupled to a PA300E preamplifier. While the MDN55EV1PA variety utilizes a straight cable connection at the back of the detector, the MDNR55EV1PA utilizes a right angle connection for use in tight spaces. These factory matched and calibrated detector pairs simplify deployment in new or existing systems by eliminating the need for calibration/matching during installation. See ordering information for details.

The MDN55EV1/MDNR55EV1 high sensitivity extended range gamma scintillation detector consists of one NaI(Tl) gamma scintillator optically coupled to a photomultiplier (PM) tube via a Lucite light pipe. A reference Light Emitting Diode (LED) and a temperature sensor are embedded in the Lucite pipe.

This detector operates with a unique gain stabilization circuitry for temperature- compensated drift-free operation, resulting in improved accuracy and extended calibration cycles.

MDN55EV1 operates in conjunction with a preamplifier, Model PA300E. The preamplifier contains circuitry that compares the LED pulses with a reference to provide closed loop feedback for automatic gain compensation. The PA300E also amplifies and shapes the detector output signal in order to provide digital pulses to the remotely located ratemeter. Procurement of matched detector/preamplifier set is recommended to simplify field maintenance and system support.

Together with its preamplifier, the MDN55EV1/ MDNR55EV1 can be connected to multifunction control and display units such as iR7040, ADM606 and ADM616.

MDN55EV1PA/MDNR55EV1PA High Sensitivity, Extended Range Scintillation Detector/Preamplifier Set

Features

  • Automatic gain stabilization using built in LED
  • Operates in pulse mode and extended range current mode
  • Linear over 9 decades
  • Calibration cycles of up to two years
  • Detectors are packaged in a water resistant O-ring sealed cylinder with a
    mu-metal shield
  • Operated with PA300E preamplifier with three single channel analyzer channels
MDN55EV2PA/MDNR55EV2PA Extended Range Gamma Scintillation Detector/Preamplifier Set

MDN55EV2PA/MDNR55EV2PA Extended Range Gamma Scintillation Detector/Preamplifier Set

Description

The MDN55EV2PA/MDNR55EV2PA matched pairs consist of a gamma scintillation detector coupled to a PA300E preamplifier. While the MDN55EV2PA variety utilizes a straight cable connection at the back of the detector, the MDNR55EV2PA utilizes a right angle connection for use in tight spaces. These factory matched and calibrated detector pairs simplify deployment in new or existing systems by eliminating the need for calibration /matching during installation. See ordering information for details.

The MDN55EV2/MDNR55EV2 high sensitivity extended range gamma scintillation detector consists of one NaI(Tl) gamma scintillator optically coupled to a photomultiplier (PM) tube via a Lucite light pipe. A reference Light Emitting Diode (LED) and a temperature sensor are embedded in the Lucite pipe.

This detector operates with a unique gain stabilization circuitry for temperature- compensated drift-free operation, resulting in improved accuracy and extended calibration cycles.

This MDN55EV2 operates in conjunction with a preamplifier, Model PA300E. The preamplifier contains circuitry that compares the LED pulses with a reference to provide closed loop feedback for automatic gain compensation. The PA300E also amplifies and shapes the detector output signal in order to provide digital pulses to the remotely located ratemeter. Procurement of matched and calibrated detector/preamplifier set is recommended to simplify maintenance and system support.

Together with its preamplifier, the MDN55EV2 /MDNR55EV2 can be connected to multifunction control and display units such as iR7040, ADM606 and ADM616.

MDN55EV2PA /MDNR55EV2PA 
Extended Range Gamma Scintillation 
Detector/ Preamplifier Set

Features

  • Automatic gain stabilization using built in LED
  • Operates in pulse mode and extended range current mode
  • Linear over 9 decades
  • Calibration cycles of up to two years
  • Detectors are packaged in a water resistant O-ring sealed cylinder with a mu-metal shield
  • Includes calibrated and matched pair PA300E preamplifier with three single channel analyzer channels
MDN455V6PA/MDNR455V6PA - Beta/Gamma Scintillation Detector/Preamplifier Set

MDN455V6PA/MDNR455V6PA Beta/Gamma Scintillation Detector/Preamplifier Set

Description

The MDN455V6PA/MDNR455V6PA matched pairs consist of a beta/gamma scintillation detector coupled to a PA300E preamplifier. While the MDN455V6PA model utilizes a straight cable connection at the back of the detector, the MDNR455V6PA utilizes a right angle connection for use in tight spaces. These factory matched and calibrated detector pairs simplify deployment in new or existing systems by eliminating the need for calibration/matching during installation. See ordering information for details.

The MDN455V6/MDNR455V6 beta/gamma scintillation detector is composed of two scintillation crystals (BGO for gamma and plastic scintillator for beta) optically coupled to a photomultiplier (PM) tube via a Lucite light pipe.

A reference Light Emitting Diode (LED) and a temperature sensor are imbedded in the Lucite pipe. This detector operates with a unique gain stabilization circuitry for temperature- compensated drift-free operation, resulting in improved accuracy and extended calibration cycles.

The MDN455V6 detector operates in conjunction with a preamplifier, Model PA300E. The preamplifier contains circuitry that compares the LED pulses with a reference to provide closed loop feedback for automatic gain compensation. The PA300E amplifies and shapes the detector output signal in order to provide digital pulses to the remotely located ratemeter. The PA300E also separates gamma and beta pulses from the composite signal by pulse shape discrimination. Procurement of matched detector/preamplifier set is recommended to simplify field maintenance and system support.

Together with its preamplifier, the MDN455V6/ MDNR455V6 can be connected to multifunction control and display unit such as iR7040, ADM606 and ADM616.

MDN455V6PA/MDNR455V6PA Beta/Gamma Scintillation Detector/Preamplifier Set

Features

  • Automatic gain stabilization using built in LED
  • Linear over 5+ decades
  • Separate crystals for beta and gamma channels
  • Calibration cycles of up to two years
  • Detectors are packaged in a water resistant O-ring sealed cylinder with a
    mu-metal shield
  • Operated with PA300E preamplifier with three single channel analyzer channels